A Validated HPLC Ketotifen Fumarate Assay Method for Cleaning Validation on an Automatic Packaging Machine
Tadakazu Tokumura, Ai Yasumoto, Takuro Kurita
Published: Nov. 30, 2018 |
245
160
DOI: 10.21276/sajp.2018.7.11.1
Pages: 460-463
Downloads
Abstract
Automatic packaging machines are used for preparing one-dose packages with
powders, granules, tablets and capsules in pharmacies in Japan. The packaging machines
are not dedicated to an individual patient, which leads to contamination of the packaging
for the next patient. Cleaning validation for pharmaceutical manufacturing plants is
therefore considered essential for packaging machines. The purpose of the present study
was to develop and validate an HPLC method for assaying ketotifen fumarate (KTF) for
use as KTF cleaning validation on an automatic packaging machine. A chromatographic
system comprising a YMC AM12S05-1506WT column, mobile phase of
CH3CN:H2O:HClO4:NaClO4=400:600:1:5 (V/V/V/W), flow rate of 1 mL/min, and UV
detector set at 300 nm was used. Propyl parahydroxybenzoate (PPB) was used as an
internal standard. The KTF and PPB retention times were approximately 6.4 and 10.8
min, respectively. Regression analysis found that the method was linear over the standard
curve range from 0.1 to 100 μg/tube. Inter-day precision and accuracy ranged between
1.70 and 22.80%, and-4.37 and 6.50%, respectively. The precision and accuracy values
were under 10% and inside a range of -10% to 10% without 0.1 μg/tube. Therefore,